#! $Id: H_2732.scr,v 1.2 2015/11/04 22:59:39 elehman Exp $ #! $Name: $ #! Description: Combination detune, obs and cal mode PZT flats, reduced focus sweep in mod-L mode #! Group: MACRO #! Wavelength ID=0xff00, Polarization ID=410, Pixel Mask=20 #! # # The cadence must be changed to 135 seconds in order to run # the original detune sequence (FTS 3027) /INT_SS_WAIT CADENCE=45 /INT_SS_WAIT SECONDS=5 /HMI_SQ_FTS_OP OP=ACTIVATE FTS_ID=3027 PRIORITY=1 /HMI_SQ_CONFIG PARAMETER=CADENCE VALUE=135 /HMI_SQ_CONFIG PARAMETER=SET_UPDATE_FLAG /INT_SS_WAIT SECONDS=120 /HMI_SQ_FTS_OP OP=CLEAR FTS_ID=3027 PRIORITY=1 /INT_SS_WAIT SECONDS=5 # The cadence must be changed back to 90 seconds /HMI_SQ_CONFIG PARAMETER=CADENCE VALUE=90 /HMI_SQ_CONFIG PARAMETER=SET_UPDATE_FLAG # # Obs mode flat # wait 70 seconds (start 20 seconds before cadence boundary) /INT_SS_WAIT CADENCE=90 /INT_SS_WAIT SECONDS=70 /HMI_SQ_FTS_OP OP=ACTIVATE FTS_ID=3041 PRIORITY=1 /HMI_RUN SCRIPT=6020 /HMI_SQ_FTS_OP OP=CLEAR FTS_ID=3041 PRIORITY=1 # # Cal mode flat # wait 25 seconds between flat fields (start 20 seconds before cadence boundary) /INT_SS_WAIT SECONDS=25 /HMI_SQ_FTS_OP OP=ACTIVATE FTS_ID=3042 PRIORITY=1 /HMI_RUN SCRIPT=6022 /HMI_SQ_FTS_OP OP=CLEAR FTS_ID=3042 PRIORITY=1 /INT_SS_WAIT SECONDS=5 # # Reduced focus sweep /HMI_SQ_FTS_OP OP=ACTIVATE FTS_ID=3040 PRIORITY=1 /INT_SS_WAIT CADENCE=90 /INT_SS_WAIT SECONDS=5 /HMI_SQ_FTS_OP OP=CLEAR FTS_ID=3040 PRIORITY=1